Sign In | Join Free | My spintoband.com |
|
Brand Name : Simens
Model Number : Simens S1000
Price : Negotiable
Payment Terms : T/T,Western Union
Delivery Time : 2-4 work days
Packaging Details : Neutral Package
Category : cow scanner
Description : digital radiography detectors
System : function of ultrasound machine
Warranty : 60 days after fixed
Condition : Siemens S1000 healthcare equipment supplies
Purpose : ultrasound
Siemens S1000 color super machine, the boot into the ultrasonic working interface is normal, the heart probe is used normally, but the superficial probe and the abdominal probe can only be used normally in the 0B mode, can not be used in other modes, the machine has a card screen phenomenon at the same time.
The signal processing and control module of S1000 color supercomputer is mainly composed of probe interface module, input and output module and electronic system module. The electronic system module includes TR, CB, RC, BE, VI and RM board, TR, RC board and probe interface module ( The TI Module) performs the ultrasonic signal processing functions together, collectively referred to as the front end portion; the BE, VI, and RM boards are responsible for the subsequent processing of the ultrasonic signal data and the overall control functions of the machine, collectively referred to as the back end portion. The power supply consists of four sections, an AC AC power box, a PSA analgo power module, a PSD digital power module, and a QuickStart Battery.
Firstly, if the system software error occurs, the fault occurs. After the system is reinstalled, the fault still exists, and the system software fault may be eliminated. The problem of RC and BE is initially judged from the hardware function. As the real-time control board of the front end, the RC board is also responsible for the control function of ultrasonic transmission and reception. It is also responsible for analgo-to-digital conversion of the 192 channels of received data of the TR board, thereby converting the echo signals into digital signals, which is convenient for the back-end software. Further processing of the technology. A failure of the RC board will cause all other functions of the machine to be abnormal. The BE board receives the digitized echo signals from the front RC board and processes them according to the operator's requirements. The failure of the BE board often causes one of the functions of the color Doppler to be abnormal.
The PSA and TI can be considered by the hardware replacement exclusion method to eliminate the BE and RC failures. The TI board can support 3 probes at the same time and can be switched at will, but the probe hot swap is not supported. When the TI board fails, it will cause the 2D area to have no image, card screen or boot code "US_IMG_17" and other error codes. The PSA is responsible for the front end power supply and provides the probe working voltage. If the PSA fault will cause the probe to be caused by the abnormal voltage supply. The function is not working properly.
Replace the PSA power supply, all the functions of the machine are used normally, and the fault is eliminated.
Repair List of Featured Ultrasonic Board & Assy | |||
Brand | System | Description | Part Number |
GE | Logiq P5 | CPU mainboard | 5168431 |
GE | Logiq P6 | CPU mainboard | |
GE | Logiq E9/vivid E9 | GFI2 | 5161631 |
GE | Voluson E6/Voluson E8 | RSR | KTI301394-2/KTI196357 |
GE | Voluson E6/Voluson E8 | RFI/ RFI21b Board | KTI300614/KTI302197-6 |
GE | Voluson S6/S8/P8 | BF64 | 5396937-2 |
GE | Voluson S6/S8/P8 | RFS board | 5364098-2/5364098-3 |
GE | Voluson S6/S10/P8 | BF192 board | 5357234 |
Aloka | Alpha 5 | mainboard | EP493700DD |
Aloka | Alpha 6 | RX Beam Former | EP555501AA |
Aloka | Alpha 6 | CW board | EP555700 |
Aloka | Alpha 6 | CPU mainboard | EP537000/EP563000 |
Aloka | Alpha 6 | TX/RX | EP555300 |
Aloka | Alpha 6/Alpha 7 | DOM | EP550000AA |
Aloka | Alpha 10 | RX Beamformer board | EP531800CE |
Aloka | Alpha 10 | wave generator Board | EP495000HH/EP495000HJ/EP495000FF |
Hitachi | F37 | TX/RS | EP557400/EP557300 |
Hitachi | F37 | mainboard | EP575700BC/EP560800 |
Aloka | F75 | USP board | EP556600 |
Aloka | F75 | Cell Board | EP558800 |
Hitachi | HI VISION Avius | RDBF board | 7513907A/7513612 |
Hitachi | Arietta 60 | CELL | EU-9160C |
Hitachi | Arietta 70 | TX | EP572300AA |
Hitachi | Arietta 70 | CELL | |
Hitachi | Noblus | CELL | |
Hitachi | Noblus | CONT-B | |
Hitachi | Ascendus | CELL | 7352830A |
Siemens | X150/G40 | BE | 10131804 |
Siemens | X300 | BE 2.0/BE 3.0/BE 7.5 | 10131990 |
Siemens | X700 | BE | 10136465 |
Siemens | X700 | RC | 10136467 |
Siemens | X700 | TI | 11014292 |
Siemens | Antares | BE | 7472421 |
Siemens | Antares | CB | 7288504 |
Siemens | S2000 | RM300 | 10852163 |
Siemens | S2000 high version/S3000 | RM301 | 10854749 |
iU22/iE33 | EMB excelsior motherboard | 453561419431/453561382231/453561419431 | |
iU22/iE33 | UMB Unified Motherboard | 453561254492/453561360251 | |
HD15 | ACB acquisition control board | 453561197145/453561197144 | |
HD15 | AIM board | 4.53561E+11 | |
CX50 | control panel | 4.53562E+11 | |
CX50 | mainboard | 453561368033/453561622163/453561495832 | |
Affiniti 70 | ACQ acquisition control board | ||
EPIQ 5/EPIQ 7 | ACQ Acquisition Module board | 453561734844/453561704246 | |
Biosound | Mylab Series | CPU mainboard | 9501361000 |
Toshiba | Aplio 300/400/500 | mainboard | PM30-38696 |
Toshiba | Aplio 500 | RX | ywm2056*a |
Toshiba | Xario 200 | TX | PM30-32732-1 |
Toshiba | Xario 200 | RX | PM30-32733-1 |
Remarks | |||
1.The repair cycle is within 3-5 days. | |||
2.Warranty after fixed:60 days. | |||
3. Kindly send us the estimated main faults (better to have image) for our engineer to prejudge. | |||
4.Package price is available if repair 3 pcs or more at one time. | |||
5.Extra discount for VIP partners. | |||
6.You could have different part No.under same description. |
![]() |
Cow Scanner Ultrasound Machine Repair Siemens S1000 Cardiac Transducer Maintenance Images |